IRAF help page for package noao.twodspec.longslit, program fluxcalib

from NOAO fluxcalib -- Apply flux calibrationUSAGEPARAMETERSDESCRIPTIONEXAMPLESREVISIONSSEE ALSO

fluxcalib -- Apply flux calibration


USAGE

fluxcalib images fluxfile


PARAMETERS

input

List of input images to be flux calibrated.

output

List of output flux calibrated images. The output images may be the same as the input images. The output image will be of type real regardless of the input pixel type.

fluxfile

Flux calibration file from onedspec.sensfunc.

fnu = no

Convert the flux calibration to flux per unit frequency (F-nu)?

exposure = otime

Exposure time keyword in image headers.


DESCRIPTION

The specified images are flux calibrated using a flux calibration image file derived from the onedspec package using standard stars. The flux calibration pixel values are in magnitudes and the pixel coordinates are in wavelength. The multiplicative calibration factor is given by the formula

factor = 10 ** (-0.4 * calibration) / exposure / dispersion.

Since the calibration data has units of (instrumental intensity) / (ergs/cm**2), the exposure time for the image must be in seconds and the pixel dispersion in wavelength/pixel to yield units of ergs/cm**2/sec/wavelength.

The calibration wavelengths are interpolated to the wavelengths of the image pixels and the correction applied to the pixel values. Note that the image pixel values are modified.

If flux per unit frequency is requested then the flux values are multiplied by

wavelength ** 2 / velocity of light (in Angstroms/sec)

to yield units of ergs/cm**2/Hz/sec/(wavelength/Angstrom). Note that normally the wavelength units should be Angstroms.

It is possible to flux calibrate images which are binned in logarithmic wavelength intervals. The point to note is that the units of the flux calibrated image will be the same. Therefore, rebinning to linear wavelength coordinates requires only interpolation and not flux conservation. When extracting standard stars from logarithmicly bin spectra for determination of a flux calibration it is necessary to rebin the extracted one dimensional spectra to linear wavelength (required by onedspec) conserving flux so that the instrumental counts are preserved.

The image header keyword DISPAXIS must be present with a value of 1 for dispersion parallel to the lines (varying with the column coordinate) or 2 for dispersion parallel to the columns (varying with line coordinate). This parameter may be added using hedit.


EXAMPLES

Standard stars were observed and extracted to one dimensional spectra. The standard stars are then used to determine a flux calibration using the onedspec package. A set of dispersion and extinction corrected images is flux calibrated in-place with the command

	cl> fluxcalib img* img* sens.0000

where "sens.0000" is the calibration file produced by the task onedspec.sensfunc.

To keep the uncalibrated image:

	cl> fluxcalib n1ext.004 n1extf.004 sens.0000

3. If the DISPAXIS keyword is missing and the dispersion is running vertically (varying with the image lines):

	cl> hedit *.imh dispaxis 2 add+

REVISIONS

REVISIONS

V2.10

The output pixel type is now forced to be real.


SEE ALSO

onedspec.standard onedspec.sensfunc,


This page automatically generated from the iraf .hlp file. If you would like your local iraf package .hlp files converted into HTML please contact Dave Mills at NOAO.

dmills@noao.edu